Z-0737-2008: Sechrist Hyperbaric Chamber Gurney, Standard Gurney Without Siderails & Stretcher Model, P/N 21487, Sechr

Recall record Z-0737-2008 for Sechrist Industries Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0737-2008
Company in FDA record
Sechrist Industries Inc
FDA classification date
Jan 18, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Brake Problems: The stretcher brake cams may crack while attempting to lock the stretcher brake (Sechrist is supplied the stretchers by the manufacturer Stryker Medical)

Source facts

Action
Urgent Medical Device Correction letters were sent to customers beginning on November 9, 2007, via FedEx & DHL overnight. Customers were notified this action was being taken to providing notification to their customers as a continued implementation of the Stryker Medical field correction. The Notification letter instructed their customers to: 1. Locate the gurneys with the serial numbers listed in this advisory. 2. If any of the affected gurneys cannot be located, determine if they have been sold, transferred, or donated to another facility and notify us regarding their new location. 3. Verify the brakes are working properly. The brake function may be checked by engaging the brake pedal and pushing on each end of the gurney to make sure that casters are locked securely. 4. Remove the gurneys with non-functioning brakes from service until brake cams have been replaced and brakes are fully functional. 5. Contract Sechrist for the replacement parts package. 6. Complete the retrofit of the Gurney. 7. Verify correct function of the brake assembly following retrofit. 8. Complete and send the Correction Verification Form contained in the package to Sechrist.
Initiated
2007-10-31
Root cause
Device Design
Terminated
2012-04-19
Distribution
Worldwide-USA, Brazil, Bulgaria, Canada, Dubai, U.A.E., Estonia, India, Indonesia, Japan, Malaysia, Saudi Arabia, Sweden, Taiwan, Thailand, and Vietnam.
Recall number
Z-0737-2008
Recall status
Terminated
CDRH record ID
67316
FDA classified
2008-01-18
Recall event number
45906

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗