Z-0737-2012: NicVue Software version 2.9.2 and 3.0.1. Medical Device Data System (MDDS) Intended use: The NicVue sof

Recall record Z-0737-2012 for CareFusion 209 Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0737-2012
Company in FDA record
CareFusion 209 Inc.
FDA classification date
Jan 13, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Under certain work flow processes, when selecting a new patient name in NicVue, the associated exams list may not properly refresh, and may continue to display the previous patient's exams.

Source facts

Action
The firm, CareFusion, sent an "URGENT: MEDICAL DEVICE CORRECTION" letter dated December 16, 2011 to its customers. The letter included a "Description of the issue" and "Actions you must take to correct this issue". The customers were instructed to follow the instructions included with the CD to update all of your NicVue systems; complete and return the SOFTWARE UPDATE VERIFICATION FORM to CareFusion-NeuroCare via fax to: +1 608 829 8517 ; mail to ATTN: Regulatory Affairs, CareFusion-NeuroCare, 1850 Deming Way, Middleton, WI USA 53562 and/or email (PDF) to: Susan.Niesen@CareFusion.com, and if you have transferred any of these products to another location, send those users a copy of this Field Correction notification and notify CareFusion-NeuroCare at +1 800 356 0007 or +1 608 829 8500 ext. 5129, or via email at susan.niesen@carefusion.com of this transfer. If assistance is needed in completing these actions, contact CareFusion NeuroCare Call Center at +1 800 356 0007 or +1 608 829 8500 for further directions; press option 2 for Technical Support.
Initiated
2011-12-16
Root cause
Software design
Terminated
2012-10-11
Distribution
Worldwide distribution: USA (nationwide) including states of: AL, AK, AZ, AR, CA, CO, DE, DC, FL, GA, HI, ID, IL, IN, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, PR, SC, SD, TN, TX, VA, WA,WV, and WI; and countries of: ARGENTINA, AUSTRALIA, BULGARIA, BANGLADESH, BELGIUM, CHILE COLOMBIA, CYPRUS, CZECH REPUBLIC, CANADA, CHINA, CROATIA, CURACO, DENMARK, DOMINICANA, EGYPT, FRANCE,GERMANY, HUNGARY, HONG KONG, IRELAND, INDIA, ITALY, JAPAN, KOREA, KUWAIT, LEBANON, LITHUANIA, MALAYSIA, MEXICO, NORWAY, NETHERLANDS, OMAN, PARAGUAY, PERU, POLAND, PORTUGAL, PALESTINIAN, PAKISTAN, PHILIPPINES,QATAR, ROMANIA, SAUDI ARABIA, SINGAPORE, SERBIA, SLOVENIA, SPAIN, SLOVAKIA, SWITZERLAND, SRI LANKA, RUSSIA, TAIWAN, THAILAND, TURKEY, UNITED KINGDOM, URUGUAY, UNITED ARB EMIRATES, and VIET NAM.
Recall number
Z-0737-2012
Recall status
Terminated
CDRH record ID
106489
FDA classified
2012-01-13
Recall event number
60797

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗