- Recall classification
- Class II
Reason stated in the source
Due to an incorrect configuration setting in the EPX database of StentBoost Live, acquisition does not stop automatically as it should. When the user selects the StentBoost Live protocol on the Philips Azurion system, the following on-screen message is displayed to the user in the Philips StentBoost Live application: "Press the cine pedal until the acquisition stops". However, due to the incorrect configuration of the EPX database, the acquisition does not stop after 40 images, but instead it continues as long as the pedal is pressed.
Source facts
- Action
- Philips notified customers of the issue via an Electronic Product Radiation Defect notification letter dated February 2, 2022. Philips will schedule an appointment with customers to install the software update. Philips will correct the affected systems at no cost to customers with a software upgrade. For information contact your local Philips representative: 1-800-722-9377.
- Initiated
- 2022-02-02
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2024-11-05
- Distribution
- US Nationwide distribution.
- Recall number
- Z-0737-2022
- Recall status
- Terminated
- CDRH record ID
- 192351
- FDA record created
- 2022-04-05
- Recall event number
- 89750