Z-0738-2019: VentStar Basic (P) 180, Catalog Number MP00351 Breathing Circuits/Anesthesia Sets without a Water trap:

Recall record Z-0738-2019 for Draeger Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0738-2019
Company in FDA record
Draeger Medical, Inc.
FDA classification date
Jan 24, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Incorrectly pre-assembled breathing circuit has potential for incorrect connection/short circuit of the breathing circuit

Source facts

Action
On Friday 12/21/2018, Draeger, Inc. issued an "Urgent Medical Device Recall Notification" and Customer Reply and Order Form to consignees via Certified Mail. Two Recall Notifications/Customer Reply and Order forms will be sent to each facility. One will be addressed to Risk Manager. The other will be addressed to Purchasing/Materials Manager. Based on customer response, Draeger will replace any affected Breathing Circuits/Anesthesia Sets Free of Charge. Customers are advised to take the following action: - Before each use, and also after any temporary disconnection, please check whether the breathing circuit is assembled such that both the inspiratory and the expiratory connector of the ventilator or anesthesia machine are each connected to the y-piece. - Please inspect your existing stock by following the inspection instructions in the attachment and complete and return the attached Customer Reply and Order form via fax to:215-372-2940 or email to:denise.frey@draeger.com to confirm to us that you have carried out the inspections. The form is also used to order Free of Charge replacements if you find any breathing circuits that have been incorrectly pre-assembled. - The correct set-up of the breathing circuit is illustrated in the Instructions for Use not only of the hose systems but also of the anesthesia machines and ventilators. Connection to the device must be checked prior to use. An incorrectly pre-assembled breathing circuit can be detected because it will not be a single cohesive hose system but two separate sub-systems. Customers with questions, please call 1-800-437-2437 (press 1 at the prompt, then 32349).between the hours of 8:00 AM 4:30 PM EST.
Initiated
2018-12-21
Root cause
Process design
Terminated
2019-04-16
Distribution
US Distribution to states of: AL, CA, CO, CT, FL, KY, MI, NJ, OH, PA, SC & TX.
Recall number
Z-0738-2019
Recall status
Terminated
CDRH record ID
170178
FDA classified
2019-01-24
Recall event number
81874

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗