Customer reprocessing methods for the reuseable hollow fiber dialyzers may damage the hollow fibers and result in blood leaks during dialysis.
Source facts
Action
Asahi visited each of the U.S. customers starting on 10/1/01, observing how the facility handles the dialyzer before pre-cleaning, during pre-cleaning and priming. The users were retrained in the reprocessing steps, emphasizing the controls and care to be taken during pre-cleaning, including the verification of water pressure.
Initiated
2001-10-01
Root cause
Other
Terminated
2004-03-24
Distribution
New Jersey, Ohio, Texas, California, Georgia, Pennsylvania, Illinois and Indiana
Recall number
Z-0739-04
Recall status
Terminated
CDRH record ID
31622
FDA classified
2004-07-20
Recall event number
28341