Philips Medical Systems North America Co. Phillips
FDA classification date
Jun 14, 2007
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
The Collimator field-defining light beam does not meet the light field contrast ratio requirement of 4:1. System is non-compliant with performance standard (1020.31 (d)(2)(iii) 21 CFR
Source facts
Action
Firm issued letter on 03/20/2007 to consignees advising them that firm representatives will contact consignee to replace collimators at no charge.
Initiated
2007-03-15
Root cause
Other
Terminated
2010-11-15
Distribution
Nationwide; Units were distributed to hospitals, radiologists, and health care clinics in IA, LA, ND, PA, and TX, and Washington DC.
Recall number
Z-0739-2007
Recall status
Terminated
CDRH record ID
51483
FDA classified
2007-06-14
Recall event number
37767