Skip to content

Public FDA record

Z-0739-2022: Nextra CH Cannulated Hammertoe Mini Instrument Kit-indicated for small bone reconstruction limited to int

Recall record Z-0739-2022 for Nextremity Solutions. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0739-2022
Company in FDA record
Nextremity Solutions
FDA record creation date
Mar 9, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During insertion of the middle phalanx implant into bone, the driver may deform the implant locking mechanism and result in inadequate mating force between the middle and proximal phalanx implants, resulting in surgical delay, in-situ removal of the implant, use of alternative surgical approach, disassembly of the implants post-operatively, and revision

Source facts

Action
Nextremity Solution issued URGENT Medical Device Recall dated 1/13/22. Letter states reason for recall, health risk and action to take: 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product (see Table 1 above) in your trunk stock inventory. 3. Immediately return all affected product from your trunk stock inventory. a. Complete Attachment 1- Medical Device Recall Return Response and send to QA@nextremity.com within three (3) days. This form must be returned even if you do not have affected products available to return in your territory. b. For each return, send a copy of Attachment 1 to QA@nextremity.com. c. Include a hardcopy of Attachment 1 in the package/box with your return shipment. d. Mark "RECALL" on the outside of the package/box. e. Ship to the address provided in Attachment 1- Medical Device Recall Return Response. 4. Retain a copy of your shipping documentation and Attachment 1- Medical Device Recall Return Response for your records in the event of a compliance audit of your facility. 5. If you have further questions or concerns after reviewing this notice, please call Nextremity Solutions at 574-376-2404 between 8:00 am and 5:00pm EST, Monday through Friday. Request to speak to a representative from quality and regulatory. Calls received outside of operating hours will receive a voicemail prompt. Alternatively, your questions may be emailed to QA@nextremity.com.
Initiated
2022-01-13
Root cause
Device Design
Terminated
2024-05-24
Distribution
US Nationwide distribution in the states of AZ, IL, GA, LA, MD, NY, PA, TX, WA.
Recall number
Z-0739-2022
Recall status
Terminated
CDRH record ID
191983
FDA record created
2022-03-09
Recall event number
89588

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗