Open original FDA record ↗Recall record
- FDA identifier
- Z-0740-2013
- Company in FDA record
- Cordis Corporation
- FDA classification date
- Jan 25, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Affected lots of Cordis ExoSeal (TM) VCD have been sterilized using a process that did not achieve the sterility assurance level that is required by the standards of this class of product. Model Numbers EX500, EX600, and EX700.
Source facts
- Action
- Cordis Corporation sent an "Urgent Medical Device Recall (Removal ) letter dated October 29, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to immediately identify and set aside all product listed in a manner that ensures the affected product will not be used. A product recall response form was attached for customers to review, complete, sign and return the enclosed acknowledgement form directly to Cordis. For further questions please call (786) 313-2000.
- Initiated
- 2012-10-29
- Root cause
- Material/Component Contamination
- Terminated
- 2013-11-06
- Distribution
- US (nationwide) including the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and Hawaii.
- Recall number
- Z-0740-2013
- Recall status
- Terminated
- CDRH record ID
- 114845
- FDA classified
- 2013-01-25
- Recall event number
- 63827