Skip to content

Public FDA record

Z-0740-2023: Abbott Alinity m Resp-4-Plex Amp Kit, Part Number 09N79-090

Recall record Z-0740-2023 for Abbott Molecular, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0740-2023
Company in FDA record
Abbott Molecular, Inc.
FDA classification date
Dec 20, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Reports of an increase in reactive negative controls and false positive results with certain lots of Alinity M Resp-4-Plex AMP kits.

Source facts

Action
Abbott Molecular issue an Urgent Field Safety Notice /Field Correction Recall to its consignees on 11/22/2022 by FedEx priority express. The notice explained the issue and requested the following actions: "- Discard inventory of any lots listed in Appendix A below. Contact Abbott Customer Support for credit of any unused kit(s). - If you have forwarded any kits of these lots to other laboratories, please inform them of this Urgent Field Safety Notice /Field Correction Recall, and provide a copy of this letter. - Return the associated Customer Reply form. Please review this information with laboratory personnel and retain this communication for future reference. If you have any questions regarding this communication, please contact your local Abbott representative, or call Abbott Technical Support at 1-800-55-7042 Option 2."
Initiated
2022-11-22
Root cause
Under Investigation by firm
Distribution
Worldwide
Recall number
Z-0740-2023
Recall status
Open, Classified
CDRH record ID
197252
FDA classified
2022-12-20
Recall event number
91202

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗