Open original FDA record ↗Recall record
- FDA identifier
- Z-0741-2012
- Company in FDA record
- Advanced Sterilization Products
- FDA classification date
- Jan 13, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
ASP determined that an internal filter in the vacuum pump on the STERRAD 200 System is subject to wear over time.
Source facts
- Action
- ASP will send an Urgent Medical Device Correction letter to all affected customers on January 23rd, 2012. The letter describes the product, problem and the actions to be taken. The letter informs customers that ASP has determined that the vacuum pump in the STERRAD 200 System may cause the emission of a slightly increased level of hydrogen peroxide (H2O2) vapor into the surrounding environment if not regularly serviced or replaced. The letter instructed customers to read the "Issue Description" and "Recommendations" sections in the letter, pass on the notice with any staff that work with or around the STERRAD 200, and maintain awareness of this communication until ASP has serviced the system. Also, a list of FAQ was provided. The letter states that ASP will replace, free of charge, the vacuum pumps on all STERRAD 200 Systems that have exceeded the pump manufacturer's recommended service interval. ASP has contracted with Stericycle to manage this field correction. Telephone numbers are provided in the letter: for questions related to this field correction, please contact Stericycle at 1-877-225-9750 and to report any complaints or suspected problems with the System, please contact ASP Professional Services directly at 1-888-783-7723, option # 2.
- Initiated
- 2012-01-23
- Root cause
- Nonconforming Material/Component
- Terminated
- 2013-09-23
- Distribution
- Worldwide Distribution - USA (nationwide)
- Recall number
- Z-0741-2012
- Recall status
- Terminated
- CDRH record ID
- 106716
- FDA classified
- 2012-01-13
- Recall event number
- 60889