Skip to content

Public FDA record

Z-0741-2023: InPen Diabetes Management App, part of the InPen System, REF: MMT-105ELBLNA, MMT-105ELGYNA, MMT-105ELPKNA

Recall record Z-0741-2023 for Companion Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0741-2023
Company in FDA record
Companion Medical, Inc.
FDA classification date
Dec 20, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Diabetes management app may crash immediately upon opening, while using the app and/or while app is running in the background, which may lead to inability to use the app. An error message may be displayed indicating the app has stopped or isn't responding. After the crash, a present dose reminder is not displayed, which may lead to delayed insulin therapy and potentially resulting in hyperglycemia.

Source facts

Action
On 11/8/22, Medtronic emailed correction notices to customers who were asked to do the following: 1) Verify the running app by: a) Open the app on the device, b) From Home screen tap the "Settings" button, c) Select the "Help and Support" option, d) Under Contact Us section, note the app version listed. 2) If affected app versions are not listed, no action is necessary. 3) If affected apps are listed, uninstall the app and reinstall the version currently available in the app store (5.5.4.4) by following these steps: a) Perform an app search for InPen, b) Tap the icon to the left of the Open button, c) Tap the Uninstall button, d) Confirm the uninstall, e) Tap the Install button. 4) After reinstall and log into the InPen app all past data will remain available and visible within the app logbook. 5) Acknowledge receipt and understanding of the correction notice. The firm provided the following 24-hour technical support phone number: 1-800-646-4633, option 1.
Initiated
2022-11-08
Root cause
Software design
Distribution
US Nationwide distribution including in the states of CA, NY, MT, OH, IN, MN, PA, CT, TX, WA, TN, SD, MI, FL, OR, NC, KY, WI, MO, WV, NV, SC, AZ, AR, MA, LA, VT, GA.
Recall number
Z-0741-2023
Recall status
Open, Classified
CDRH record ID
197054
FDA classified
2022-12-20
Recall event number
91158

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗