- FDA identifier
- Z-0742-2012
- Company in FDA record
- Groman Inc.
- FDA classification date
- Jan 13, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Groman Inc. of Margate, FL recalled their product Cordless Etcher Air Brush. When non-conforming cartridges are inserted into the Cordless Etcher they can induce excessive stresses on the device causing cracks in the plastic housing of the device. These crack can lead to a hazardous situation due to the high pressure of the gas contained in the CO2 cartridges.
Source facts
- Action
- Groman Inc. sent a "HAZARD NOTICE: IMPORTANT NOTICE ABOUT CORDLESS ETCHER HAZARD DUE TO IMPROPER USE" letter dated March 1, 2011 to all affected customers. The letter describes the product, problem, and advises the customers of the proper use of the device. Customers with questions and concerns, should contact the President of Groman Dental at 800-906-9298 ext. 709.
- Initiated
- 2010-12-13
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-01-17
- Distribution
- Worldwide Distribution-USA (nationwide) and the country of Rotterdam, Netherlands.
- Recall number
- Z-0742-2012
- Recall status
- Terminated
- CDRH record ID
- 100477
- FDA classified
- 2012-01-13
- Recall event number
- 58889