Skip to content

Public FDA record

Z-0742-2012: New Label: EtchMaster ***Cordless Handpiece***P/N 797 www.GromanDental.com 1800-906-9298 HAZARD WARNING:

Recall record Z-0742-2012 for Groman Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0742-2012
Company in FDA record
Groman Inc.
FDA classification date
Jan 13, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Groman Inc. of Margate, FL recalled their product Cordless Etcher Air Brush. When non-conforming cartridges are inserted into the Cordless Etcher they can induce excessive stresses on the device causing cracks in the plastic housing of the device. These crack can lead to a hazardous situation due to the high pressure of the gas contained in the CO2 cartridges.

Source facts

Action
Groman Inc. sent a "HAZARD NOTICE: IMPORTANT NOTICE ABOUT CORDLESS ETCHER HAZARD DUE TO IMPROPER USE" letter dated March 1, 2011 to all affected customers. The letter describes the product, problem, and advises the customers of the proper use of the device. Customers with questions and concerns, should contact the President of Groman Dental at 800-906-9298 ext. 709.
Initiated
2010-12-13
Root cause
Nonconforming Material/Component
Terminated
2012-01-17
Distribution
Worldwide Distribution-USA (nationwide) and the country of Rotterdam, Netherlands.
Recall number
Z-0742-2012
Recall status
Terminated
CDRH record ID
100477
FDA classified
2012-01-13
Recall event number
58889

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗