Skip to content

Public FDA record

Z-0742-2020: Axis-Shield Liquid Stable (LS) 2-Part HOMOCYSTEINE REAGENT, Part Number FHBC100 - Product Usage: The devi

Recall record Z-0742-2020 for Axis-Shield Diagnostics, Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0742-2020
Company in FDA record
Axis-Shield Diagnostics, Ltd.
FDA record creation date
Dec 26, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential performance issue as detected by an under-recovery of sample results during routine testing.

Source facts

Action
Urgent Medical Device Correction notification letters dated 9/16/19 were sent to customers. Necessary Actions Immediately discontinue use of the affected lots listed and switch to the alternate material lot(s) you may have available in your laboratory. Immediately destroy any remaining inventory of the affected lots listed according to your laboratory procedures. Complete and return the accompanying verification form Immediately order replacement material(s) if you do not have unaffected material available. Please retain this letter for your laboratory records.
Initiated
2019-09-16
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the states of CA, KY.
Recall number
Z-0742-2020
Recall status
Open, Classified
CDRH record ID
177056
FDA record created
2019-12-26
Recall event number
84043

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗