Open original FDA record ↗Recall record
- FDA identifier
- Z-0742-2021
- Company in FDA record
- Westmed, Inc.
- FDA record creation date
- Dec 29, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a potential that the fitment between the oxygen delivery tube and the mask adapter may become loose. This could result in the device tubing not making a secure connection to the mask adapter resulting in a possible delay in treatment.
Source facts
- Action
- On November 11, 2020, West Med, Inc. issued a recall notification for oxygen delivery products via E-Mail to affected consignees. In addition to informing consignees about the recall products, the firm asked consignees to take the following actions: 1. Please complete the attached response form and quarantine affected product. 2. DO NOT DESTROY OR DISPOSE OF PRODUCT AT THIS TIME. 3. Contact Kalyn Galvez-Latneau, Westmed Customer Service Manager, for additional information and to obtain a Return Material Authorization (RMA). Kalyn Galvez-Latneau, kgalvez@westmedinc.com, (520) 294-7987, Ext 1232.
- Initiated
- 2020-11-11
- Root cause
- Process control
- Terminated
- 2023-02-07
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of PA, OK, MO, CO, RI, IL, CA, TX, OH, MD, IL, MN, KS and the countries of Canada, Korea, United Kingdom
- Recall number
- Z-0742-2021
- Recall status
- Terminated
- CDRH record ID
- 184798
- FDA record created
- 2020-12-29
- Recall event number
- 86903