Z-0742-2025: Galaxy System (GAL-001). Intended to provide bronchoscopic visualization of and access to patient airways

Recall record Z-0742-2025 for Noah Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0742-2025
Company in FDA record
Noah Medical
FDA classification date
Dec 20, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Bronchoscopic visualization system power supplies have defective semiconductor components that may fail unexpectedly, which could result in an aborted procedure since it is a non recoverable failure. Manual removal of the bronchoscope when the system is not functioning can increase the chance of injury to the tissues of the lung as the scope is withdrawn including the possibility of pneumothorax.

Source facts

Action
On 10/28/24, recall notices were sent to customers who were told the firm would contact them to arrange inspection and replacement of the power supply system. Share this notification with any other healthcare professionals who may use the system. Complete and return the response form. Firm customer Service: customerservice@noahmed.com, 888-325-6624
Initiated
2024-10-28
Root cause
Component change control
Distribution
US: SC, OH, PA, TN, MO, IN, CA, WV, NY, MD, MA, NJ, ID, CT, NC
Recall number
Z-0742-2025
Recall status
Open, Classified
CDRH record ID
210678
FDA classified
2024-12-20
Recall event number
95585

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗