Open original FDA record ↗Recall record
- FDA identifier
- Z-0742-2025
- Company in FDA record
- Noah Medical
- FDA classification date
- Dec 20, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Bronchoscopic visualization system power supplies have defective semiconductor components that may fail unexpectedly, which could result in an aborted procedure since it is a non recoverable failure. Manual removal of the bronchoscope when the system is not functioning can increase the chance of injury to the tissues of the lung as the scope is withdrawn including the possibility of pneumothorax.
Source facts
- Action
- On 10/28/24, recall notices were sent to customers who were told the firm would contact them to arrange inspection and replacement of the power supply system. Share this notification with any other healthcare professionals who may use the system. Complete and return the response form. Firm customer Service: customerservice@noahmed.com, 888-325-6624
- Initiated
- 2024-10-28
- Root cause
- Component change control
- Distribution
- US: SC, OH, PA, TN, MO, IN, CA, WV, NY, MD, MA, NJ, ID, CT, NC
- Recall number
- Z-0742-2025
- Recall status
- Open, Classified
- CDRH record ID
- 210678
- FDA classified
- 2024-12-20
- Recall event number
- 95585