Baxter's Service Center did not perform the ground impedance and leakage current test which verify if any power overflow occurs and that it is properly discharged through the device electrical system.
Source facts
Action
Baxter sent service technicians to the consignees location on July 14, 2010 and October 6, 2010 and conducted the ground impedance and leakage current test on the affected pumps. The pumps were found with specifications.
Initiated
2010-07-14
Root cause
Employee error
Terminated
2012-01-13
Distribution
Nationwide distribution: Puerto Rico.
Recall number
Z-0744-2012
Recall status
Terminated
CDRH record ID
105691
FDA classified
2012-01-13
Recall event number
60520