Skip to content

Public FDA record

Z-0744-2022: PuraSinus Absorbable Nasal Hemostat, Part Numbers 651-006

Recall record Z-0744-2022 for 3-D Matrix, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0744-2022
Company in FDA record
3-D Matrix, Inc.
FDA record creation date
Mar 10, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The boxes used by the shipper were not validated for use during transport.

Source facts

Action
A notification letter dated 12/2/19 was sent to the customer. The purpose of this letter is to inform you of the market withdrawal of 3 units of PuraSinus peptide hydrogel, lot#: 19E30A50, Expiration April 2022 (UDI #: (0l)00860001820829(17)220430(10)19E30A50) shipped to your location on October 7, 2019. The reason for this market withdrawal is that the boxes used by the shipper located in Florida are boxes that are designed to maintain proper temperatures during transport, however the boxes that were used are not the boxes that we have validated for use during transport. While 3-D Matrix Medical Technology is reasonably certain that the product was not damaged or degraded in any way - because upon arrival the contents of the package were observed to still be cool and our internal testing has demonstrated product stability over a much wider range of temperatures (product is stable from -5 C to 30 C) than our specified shipping temperatures in transit to your office, in an abundance of caution we are requesting the market withdrawal of the product from consideration for use in your patients immediately and encourage you to return the three units of PuraSinus to the manufacturer at your earliest convenience. 3-D Matrix Inc. will provide shipping instructions and a prepaid label for your use in facilitating the return in the coming days. We apologize for this inconvenience and if acceptable, 3-D Matrix, Inc. will ship replacement products in the proper shipping containers Please contact 3-D Matrix, Inc. with any questions at infous@puramatrix.com or 1-781-373-9020. As a trusted partner in healthcare, we are committed to your patient's safety.
Initiated
2019-12-02
Root cause
Packaging process control
Terminated
2023-08-25
Distribution
US Nationwide distribution in the state OH.
Recall number
Z-0744-2022
Recall status
Terminated
CDRH record ID
191969
FDA record created
2022-03-10
Recall event number
89577

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗