Siemens Healthcare Diagnostics has received complaints of IRON calibrations outside of acceptable calibration slope (m) values (0.97 - 1.03) when using IRON calibrator lots 2DN060 and 2FN063. Lots 2GN068 and 2JN074 may also be impacted but no complaints have been received.
Source facts
Action
Siemens sent a "CUSTOMER NOTIFICATION" letter dated December 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact Siemens Technical Solutions Center at 1-800-441-9250 or your local Siemens Healthcare Diagnostics representative for assistance with this issue.
Initiated
2012-12-20
Root cause
Nonconforming Material/Component
Terminated
2014-01-08
Distribution
Nationwide Distribution.
Recall number
Z-0745-2013
Recall status
Terminated
CDRH record ID
115333
FDA classified
2013-01-29
Recall event number
64081