Open original FDA record ↗Recall record
- FDA identifier
- Z-0745-2019
- Company in FDA record
- Bayer Medical Care, Inc.
- FDA record creation date
- Jan 15, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
If the head mount knob is not properly tightened and the operator directly uses the injector head to move the unit, it may become disengaged, causing the injector head to become loose or fall.
Source facts
- Action
- Urgent Medical Device Field Corrective Action notification letters were sent to customers on 12/12/18.
- Initiated
- 2018-12-14
- Root cause
- Device Design
- Terminated
- 2020-06-16
- Distribution
- The products were distributed US nationwide. The products were distributed to the following foreign countries: Algeria, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Belgium, Bermuda, Brazil, Brunei Darussalam, Canada, Chile, China, Cote D'Ivoire, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Faroe Islands, Finland, France, French Guiana, French Polynesia, Germany, Ghana, Greenland, Guam, Vatican City State, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Martinique, Mauritius, Mayotte, Mexico, Monaco, Mongolia, Morocco, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestine, Panama, Philippines, Poland, Portugal, Reunion, Russian Federation, Saint Pierre and Miquelon, Saudi Arabia, Serbia and Montenegro, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey UAE, UK, Uruguay, Venezuela, Vietnam.
- Recall number
- Z-0745-2019
- Recall status
- Terminated
- CDRH record ID
- 169990
- FDA record created
- 2019-01-15
- Recall event number
- 81808