Skip to content

Public FDA record

Z-0746-2016: Astral 100, 150 Model: 27001, 27003 In the US, Astral devices are primarily used as non-invasive homecare

Recall record Z-0746-2016 for Resmed Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0746-2016
Company in FDA record
Resmed Corporation
FDA classification date
Feb 3, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Resmed Corporation is recalling Astral 100, 150 External Battery because of a start-up issue.

Source facts

Action
ResMed sent an Urgent Safety Alert dated January 13, 2016, to all affected customers. The Alert identified the product, the problem, and the action to be taken by the customer. The notification letter provided the following instructions to distributors Field devices can be upgraded using ResMed Service Software or at an authorized ResMed Service Center. Contact Astral Support for further details. The SR2 software update will be available free of charge and should be implemented for all devices within the two-year service cycle or before at the customers preference. To acknowledge receipt of this Astral Safety Alert, submit this form. For any questions call Astral Support: 855-245-4640
Initiated
2016-01-13
Root cause
Software design
Terminated
2016-04-22
Distribution
Worldwide Distribution - US including VA, FL, MO, CA, LA, TX, NJ, NH, NC, WV, MI, ID, OK, MI, MN, UT, CT, IL, AZ, UT, KY, MT, PA, RI, TN, DE, NE, LA, NY, SC, WY, IN, VA, GA, CO and Internationally to CA, CO, and AR.
Recall number
Z-0746-2016
Recall status
Terminated
CDRH record ID
143003
FDA classified
2016-02-03
Recall event number
73071

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗