Open original FDA record ↗Recall record
- FDA identifier
- Z-0746-2017
- Company in FDA record
- Steris Corporation
- FDA classification date
- Dec 12, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The Drager Pendula spring arm device has a set of screws located on the spring arm that were not properly secured with Loctite and may become loose.
Source facts
- Action
- Steris sent an Urgent Voluntary Field Correction Notice to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to remind staff personnel that Harmony Insight Diagnostic Monitor Support Systems, including the affected Drager Pendula spring arms, are not designed or intended to be placed over hospital personnel or patients and should be moved with care. Customers with questions were instructed to call 440-392-7358 or Steris Customer Service at 1-800-548-4873. For questions regarding this recall call 440-392-7231.
- Initiated
- 2016-10-05
- Root cause
- Component design/selection
- Terminated
- 2017-04-14
- Distribution
- Nationwide Distribution including AK, AZ, CA, CO, FL, IL, KY, MI, MO, NB, NJ, NY, OH, PA, RI, SD, TN, TX, UT, WA & WV.
- Recall number
- Z-0746-2017
- Recall status
- Terminated
- CDRH record ID
- 150180
- FDA classified
- 2016-12-12
- Recall event number
- 75385