Z-0746-2020: HeartMate 3 Mobile Power Unit, REF 107754, UDI or GTIN: 00813024010883, and 107758, UDI or GTIN: 00813024

Recall record Z-0746-2020 for Thoratec Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0746-2020
Company in FDA record
Thoratec Corp.
FDA record creation date
Dec 27, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Excessive static electricity can potentially cause unrecoverable power loss and damage to the mobile power unit that can be used to power the left ventricular assist system, which could lead to no or reduced blood flow from the left ventricular assist system.

Source facts

Action
On 12/02/19, Urgent Medical Device Recall notices were delivered in person by field representatives who provided training on the methods to reduce generation of excessive static electricity. Clinician were advised the following: Recognition of alarms if the Left Ventricular Assist System (LVAS), when connected to the MPU Module, has been impacted by excessive static electricity. " If a patient is not sleeping or resting, battery power is recommended instead of the MPU Module to power the LVAS in order to reduce the impact of static electricity. " In any case of an unexpected "no external power" alarm, patients should connect to battery power to the LVAS before attempting to determine the cause of the alarm. If the alarm persists, the patient should call the Hospital Contact. " To prevent the generation of excessive static electricity during daily activities, the firm is providing clarification for static electricity management when using the MPU Module with the LVAS. Refer to Appendix A for specific reminders and recommendations to prevent excessive static electricity. Contact all patients with affected devices as soon as possible and provide them with the information in Appendix A. Also, please complete the attached acknowledgement form. Clarifications will be added to the Instructions for Use and Patient Handbook in the future. Customers with additional questions are encouraged to call 1-800-456-1477 or to contact their field representative. Using battery power will help avoid system damage from static electricity. If you're on the MPU, switch to battery power when doing things that cause static electricity. Reduce static electricity, with products such as: 1) A humidifier to add moisture to the air, 2) Dryer sheets and fabric softeners to reduce the buildup of static electricity in clothes and bedsheets, 3) Anti-static spray on carpets and other materials to reduce the buildup of static electricity 4) Skin moisturizers to make your skin less
Initiated
2019-12-02
Root cause
Device Design
Terminated
2022-09-14
Distribution
US Nationwide distribution in the states of MN, PA, IL, NJ, WI, AZ, AR, TN, MO, TX, MA, NE, MI, CA, NC, CO, OH, DE, FL, NY, IN, GA, VA, DC, CT, OK, UT, KY, MD, OR, NM, ME, SC, IA, WA, LA, NV, AL, KS, MS, WV.
Recall number
Z-0746-2020
Recall status
Terminated
CDRH record ID
178196
FDA record created
2019-12-27
Recall event number
84379

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗