Open original FDA record ↗Recall record
- FDA identifier
- Z-0747-2017
- Company in FDA record
- PerkinElmer, Inc
- FDA record creation date
- Dec 12, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Filter Tips molded incorrectly and may not seal to the Varispan tip adaptor properly causing inaccurate aspiration and dispensing
Source facts
- Action
- PerkinElmer initiated recall on June 24, 2016 to the Product Manager and was delivered to the customer by email facilitated through PerkinElmer Sales Team.Customers were requested to take the following actions until PerkinElmer's corrective measures are completed: Inspect customer inventory of the 6000689 25 Clear Filter Pipette Tip and identify if the affected lots are present. Complete and return the attached Response Form with the required information. Destroy all quantities of tips from the affected lots. Upon receipt of the return Response Form, a shipment of replacement tips will be shipped.
- Initiated
- 2016-06-24
- Root cause
- Process control
- Terminated
- 2018-07-31
- Distribution
- Worldwide Distribution - US including CA, MA, NY, IA, OR , PA, WA and Internationally to Belgium, Canada, Finland, France, Germany, Japan, Netherlands, Singapore, and UK
- Recall number
- Z-0747-2017
- Recall status
- Terminated
- CDRH record ID
- 150563
- FDA record created
- 2016-12-12
- Recall event number
- 75508