Open original FDA record ↗Recall record
- FDA identifier
- Z-0747-2022
- Company in FDA record
- Ion Beam Applications S.A.
- FDA record creation date
- Mar 10, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
When resuming from a beam pause, the Proton Therapy System does not verify whether the beam range has not been manually modified during the pause and is still the prescribed one. Therefore, if an operator manually modified the range during a pause, there is a risk that a portion of the treatment beam after the resume is delivered with an error in range
Source facts
- Action
- IBA issued Urgent Medical Device Correction notification on February 15, 2022 directly to customer either by hand or by email. Letter states reason for recall, health risk and action to take: For Proton Therapy sites where the system is operated by the customer: - Warn operators to not manually change the beam range (action called manual set range ) while a patient is on the Patient Positioning System. - IBA will provide a training to all operators about the correct use of the action called manual set range and the risk associated to its use while a patient is on the Patient Positioning System. This training will be provided on your site by August 31, 2022. Contact: Vigilance@iba-group.com. Help Desk: +32 2 507 20 81 (available 24/7)
- Initiated
- 2022-02-15
- Root cause
- Device Design
- Distribution
- IL, KS, LA, FL, MI, OK, PA, NJ, TN, TX, VA, WA Foreign: Belgium, England, Germany, France, Czech Republic, Italy, Poland, Russia, Spain, Sweden, South Korea, Taiwan, Netherlands,
- Recall number
- Z-0747-2022
- Recall status
- Open, Classified
- CDRH record ID
- 192319
- FDA record created
- 2022-03-10
- Recall event number
- 89731