Z-0747-2022: ProteusPLUS and ProteusONE- designed to produce and deliver a proton beam for the treatment of patients w

Recall record Z-0747-2022 for Ion Beam Applications S.A.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0747-2022
Company in FDA record
Ion Beam Applications S.A.
FDA record creation date
Mar 10, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When resuming from a beam pause, the Proton Therapy System does not verify whether the beam range has not been manually modified during the pause and is still the prescribed one. Therefore, if an operator manually modified the range during a pause, there is a risk that a portion of the treatment beam after the resume is delivered with an error in range

Source facts

Action
IBA issued Urgent Medical Device Correction notification on February 15, 2022 directly to customer either by hand or by email. Letter states reason for recall, health risk and action to take: For Proton Therapy sites where the system is operated by the customer: - Warn operators to not manually change the beam range (action called manual set range ) while a patient is on the Patient Positioning System. - IBA will provide a training to all operators about the correct use of the action called manual set range and the risk associated to its use while a patient is on the Patient Positioning System. This training will be provided on your site by August 31, 2022. Contact: Vigilance@iba-group.com. Help Desk: +32 2 507 20 81 (available 24/7)
Initiated
2022-02-15
Root cause
Device Design
Distribution
IL, KS, LA, FL, MI, OK, PA, NJ, TN, TX, VA, WA Foreign: Belgium, England, Germany, France, Czech Republic, Italy, Poland, Russia, Spain, Sweden, South Korea, Taiwan, Netherlands,
Recall number
Z-0747-2022
Recall status
Open, Classified
CDRH record ID
192319
FDA record created
2022-03-10
Recall event number
89731

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗