- FDA identifier
- Z-0747-2024
- Company in FDA record
- Smiths Medical Asd Inc
- FDA classification date
- Jan 17, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.
Source facts
- Action
- An URGENT MEDICAL DEVICE RECALL notification letter dated 11/10/23 was sent to customers. Required Actions for Users: 1. Please discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. 2. Inform potential users of the product in your organization of this notification and complete and return the attached response form to smithsmedical7166@sedgwick.com within ten days of receipt to acknowledge your understanding of this notification, even if you do not have the affected product and/or it has already been used. 3. If you have distributed the product further, immediately notify your accounts that receive the product identified in the Affected Items / Table 1 sections of this notification and ask them to contact Sedgwick at 1-855-215-5142 (MF, 8am-5pm ET) to obtain a response form. 4. If the product has not already been used, please return affected lots to Smiths Medical and Smiths Medical will offer replacement units. Global Complaint Management globalcomplaints@icumed.com 1-(866)-216-8806 Customer Service 1-(800)-258-5361
- Initiated
- 2023-11-10
- Root cause
- Nonconforming Material/Component
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, UAE.
- Recall number
- Z-0747-2024
- Recall status
- Open, Classified
- CDRH record ID
- 204586
- FDA classified
- 2024-01-17
- Recall event number
- 93506