Open original FDA record ↗Recall record
- FDA identifier
- Z-0748-2022
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA record creation date
- Mar 11, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The clip store function in the ultrasound imaging system does not work when the system has a disk full error. This could cause a delay in treatment if the ultrasound system is unable to save clips as study documentation during a high risk procedure, such as a stress echo exam.
Source facts
- Action
- The firm, Siemens Healthineers, sent "URGENT MEDICAL DEVICE SAFETY CORRECTION" Letters Juniper 1.0 (VA10D/VA10E/VA10F) to affected customers via Siemens update program US026/21/S informing them of the issue and advising them the firm will correct this issue with a free-of-charge software update to their ACUSON Juniper ultrasound system. The customers were instructed to do the following: Check the available storage on the ultrasound system prior to starting a study where clips are vital to the documentation, such as stress echo. If available, use a different ultrasound system capable of performing a stress echo exam until the problem is resolved. If another ultrasound system is not available, consider using other imaging modalities to achieve the diagnostic outcome. In the event that a defect is encountered, a reinstallation of the system software can resolve this problem until the available storage on the ultrasound system is again exceeded. The customers' respective customer service engineer will contact them to schedule a facility visit to update the system or inform them of a remote update. If you have further questions regarding this safety correction, please send all inquiries to: Siemens Healthineers USD Complaints usd-complaint.team@siemens-healthineers.com.
- Initiated
- 2022-01-19
- Root cause
- Software design
- Distribution
- Worldwide distribution: US (nationwide) and countries of: Algeria, Andorra, Argentina, Australia, Austria, Bangladesh, Belgium, Bosnia Herzeg, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Faroe Islands, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Kazakhstan, Kosovo, Kuwait, Latvia, Liechtenstein, Lithuania, Luxembourg, Malaysia, Mexico, Morocco, Myanmar, Namibia, Netherlands, New Zealand, Norway, Oman, Peru, Philippines, Poland, Portugal, Qatar, Rep. of Yemen, Republic Korea, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sudan, Sweden, Switzerland, Taiwan, Thailand, Turkey, U.A.E., Ukraine, United Kingdom, Uruguay, Vietnam, and Zimbabwe.
- Recall number
- Z-0748-2022
- Recall status
- Open, Classified
- CDRH record ID
- 191763
- FDA record created
- 2022-03-11
- Recall event number
- 89499