Open original FDA record ↗Recall record
- FDA identifier
- Z-0748-2024
- Company in FDA record
- Olympus Corporation of the Americas
- FDA classification date
- Jan 18, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Olympus identified inconsistencies in the Instructions for Use regarding cleaning the fiber tip and fiber cleaving and stripping instructions. Following the current instructions presents the potential for the fiber to break during use, which may cause burns, procedure delays, prolonged surgery, and foreign body in patient.
Source facts
- Action
- On December 1, 2023, the firm issued URGENT MEDICAL DEVICE CORRECTION letters to affected customers. Customers were provided with copies of the labeling addendum to add to remaining inventory. Customers may continue to use the product, but this should be done in accordance with the labeling addendum. If you require additional information, please contact Olympus at (647) 999-3203 or at Cynthia.Ow@Olympus.com.
- Initiated
- 2023-12-01
- Root cause
- Labeling Change Control
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Europe, Hong Kong, Australia, Japan, Singapore, and Latin America.
- Recall number
- Z-0748-2024
- Recall status
- Open, Classified
- CDRH record ID
- 205049
- FDA classified
- 2024-01-18
- Recall event number
- 93649