Skip to content

Public FDA record

Z-0749-2018: Vis-U-All High Temp 4"x8" Heat Seal Pouch 200 pouches per box; 5 boxes per case Product Usage: The Vis-U

Recall record Z-0749-2018 for Steris Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0749-2018
Company in FDA record
Steris Corporation
FDA record creation date
Feb 27, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

On June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch.

Source facts

Action
Steris sent an Urgent Voluntary Recall Notice letter dated July 2017 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately inspect on-hand inventory for recalled product. If you have product in your possession, please quarantine it and contact firm to coordinate shipment of replacement product. If you have questions regarding this matter, please contact STERIS Customer Service at 1-800-548-4873.
Initiated
2017-06-15
Root cause
Component change control
Terminated
2018-07-17
Distribution
US Nationwide Distribution
Recall number
Z-0749-2018
Recall status
Terminated
CDRH record ID
161348
FDA record created
2018-02-27
Recall event number
79100

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗