Open original FDA record ↗Recall record
- FDA identifier
- Z-0750-2017
- Company in FDA record
- Roche Molecular Systems, Inc.
- FDA record creation date
- Dec 13, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The LightCycler 480 algorithm used for the LightMix Zika rRT-PCR Test, EUA (catalog number 07987897001) occasionally (~1% of the time) calls a positive result for a negative sample if there is an inflection point in the fluorescence signal.
Source facts
- Action
- Roche Molecular Diagnostics Inc. sent an Urgent Medical Device Correction Letter dated November 16, 2016, to all affected customers via UPS. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use of the affected product until Roche provides updated Instructions for Use to assist in result interpretation. If product was further distributed, ensure the Urgent Medical Device Correction letter is provided to that site. Also, customers were instructed to completed the attached fax form and fax it to 1-888-912-2188. For questions customers should call 1-800-526-1247.
- Initiated
- 2016-11-16
- Root cause
- Software design
- Terminated
- 2017-12-12
- Distribution
- Nationwide Distribution
- Recall number
- Z-0750-2017
- Recall status
- Terminated
- CDRH record ID
- 151271
- FDA record created
- 2016-12-13
- Recall event number
- 75786