Open original FDA record ↗Recall record
- FDA identifier
- Z-0750-2024
- Company in FDA record
- Xcision Medical Systems, LLC
- FDA classification date
- Jan 18, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Wrong structures imported into the GammaPod Treatment Planning System.-system allowed an RTSS generated from a previous CT scan to be selected
Source facts
- Action
- Xcision issued Urgent Field Correction Letter via email on 11/30/23. Letter states reason for recall, health risk and action to take: Please post this notification internally for awareness of GammaPod BY CUSTOMER/USER Treatment Planning System users. Xcision is working on an update to the TPS software to implement controls to identify not only the patient identification but also the series of which the structures were delineated. This change is expected for implementation in early 2024. If you need any further information or support concerning this issue, please contact Xcision Customer Support at 855-924-7466 (1-855- SUPPORT XCISION).
- Initiated
- 2023-11-30
- Root cause
- Software design
- Distribution
- Worldwide - US Nationwide distribution in the states of MD, PA, TX and the country of Italy.
- Recall number
- Z-0750-2024
- Recall status
- Open, Classified
- CDRH record ID
- 205111
- FDA classified
- 2024-01-18
- Recall event number
- 93684