Skip to content

Public FDA record

Z-0750-2025: BYTE Aligner System REF PRS-0063 BYTE All-Day Aligner System BYTE AT-Night Aligner System A custom-made

Recall record Z-0750-2025 for Straight Smile, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0750-2025
Company in FDA record
Straight Smile, LLC
FDA classification date
Dec 23, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Virtual-only clinical workflow utilized for dental aligners and impression to approve patients for aligner devices may not be suitable for patients who have certain contraindications.

Source facts

Action
On 11/04/2024 and 11/07/2024, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Letter via email to customers informing them of the potential risks associated with BYTE Aligners that the patient onboarding workflow may not provide adequate assurance that patients with active periodontal disease, severe open bite, severe overjet, tooth malocclusion requiring surgical correction, mixed dentition, dental prosthetics or dental implants, or adolescents with skeletally narrow jaws do not enter treatment with BYTE Aligners. The BYTE workflow may not be adequate to identify the conditions for which BYTE Aligners are contraindicated because an in-person exam is not performed. This means that potential patients with one of the above issues may not be appropriately filtered out from treatment. For customers who are actively in treatment. To minimize the risk of illness or injury, we recommend the following: For customers who knowingly suffer from active periodontal disease, severe open bite, severe overjet, tooth malocclusion requiring surgical correction, mixed dentition, have dental prosthetics or dental implants, or are an adolescent with a skeletally narrow jaw, discontinue aligner treatment immediately and visit a dental professional. For all other customers, please promptly visit a dental professional to confirm whether continuing aligner treatment is appropriate for you. Following your dentist visit, please visit https://patient.byte.com/support to upload (i) the name and contact information for the dental professional you visited, (ii) the date visited, and (iii) any visit notes provided by the dental professional during your visit. BYTE will reach out to you to gather a signed consent to enable us to gather other information needed from such dental professional. For questions/concerns - contact treating dentist by submitting a support request through Patient Account Center (https://patient.byte.com/support), the My Byte App, or calling Byte support (877) 298-36
Initiated
2024-11-04
Root cause
Device Design
Distribution
Worldwide - U.S. Nationwide distribution including in the states of CO, KS, MI, IL, GA NY, MN, IN, PA, TX, NJ, MS, RI, SD, FL, KY, CA, OR, WI, WY, MA, NC, WA, AR, UT, OH, SC, NM, CT, TN, LA, NE, AL, MO, VA, MD, IA, OK, NV, DE, VT, AZ, RI, HI, WV, NH, MT, AK, ID, Puerto Rico, ME and the countries of Australia.
Recall number
Z-0750-2025
Recall status
Open, Classified
CDRH record ID
211093
FDA classified
2024-12-23
Recall event number
95745

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗