Skip to content

Public FDA record

Z-0751-2020: Blood Glucose Monitoring System, Model Number 08453063001 ACCU-CHEK GUIDE SAMPLE KIT FCN MG/DL US - Produ

Recall record Z-0751-2020 for Roche Diabetes Care, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0751-2020
Company in FDA record
Roche Diabetes Care, Inc.
FDA classification date
Dec 30, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential power issues with the Accu-Chek Guide Blood Glucose Monitoring System.

Source facts

Action
On October 23, 2019 the firm distributed Urgent Medical Device Recall Communication letters by mail for the Accu-Chek Guide blood glucose meters with a serial number below 92311000001. The serial number can be found on the back of the meter, where indicated by the red rectangle in the photo below. It instructs them to: 1) Always have a spare set of batteries. The meter requires two (2) size CR2032 batteries. 2) Be aware that battery life may vary due to factors such as temperature and battery manufacturer. 3) Have a back-up testing method available. 4) If you have questions regarding this notification or you need further support, please contact our Accu-Chek Customer Care team through our website (accuchek.com), email address (accu-chek.care@roche.com), or call (1-800-858-8072). 5) If there is a Business Reply Card enclosed with this mailing, please complete and return it as instructed on the Business Reply Card. 6) Keep a copy of this notification for your records. If you are currently experiencing a power issue with your Accu-Chek Guide meter, follow the instructions below. If the meter is powered on and displaying an E-9 error or low battery icon: a. Move the meter to a more moderate environment, such as ambient room temperature, to ensure optimal battery performance. b. Turn the meter off, then turn the meter back on. c. If the problem persists, perform a meter reset by doing the following: i) Remove the batteries. ii) Press and hold the Power/Set/OK button for at least 2 seconds. iii) Reinsert the batteries. d. If the problem still occurs after performing a meter reset, then replace the batteries with new ones according to the instructions in the user manual. Use high quality batteries in the meter, such as Panasonic. The meter requires two (2) size CR2032 batteries. e. If replacing the batteries does not correct the situation, please contact our Accu-Chek Customer Care team through our website (accu-chek.com), email address (accu-chek
Initiated
2019-09-30
Root cause
Device Design
Terminated
2020-07-29
Distribution
US Nationwide distribution and countries of AR, AT, AU, BE, BR, CA, CH, CL, DE, DK, EC, ES, FI, FR, GB, GR, HK, HR, ID, IN, IT, JP, KR, LB, LT, LU, MM, MY, NL, NO, PH, PK, PT, SE, SG, TH, TW, ZA.
Recall number
Z-0751-2020
Recall status
Terminated
CDRH record ID
177241
FDA classified
2019-12-30
Recall event number
84119

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗