Reason stated in the source
Due to manufacturing defects observed on Locking Caps that affect the ability for components to mate.
Source facts
- Action
- Update 12/18/2024 - The firm sent an updated Customer Notification Letter on 12/12/2024 to customers that have not responded to the initial notification on 10/29/2024. On 10/29/2024, firm sent an "URGENT: MEDICAL DEVICE RECALL" Letter to customer informing them that the firm has identified a nonconformance with the threads of the locking cap components that may be incongruous with optimal manufacturing specifications. The nonconformance can result in a locking cap that does not mate appropriately with the screw tulip. Customers are instructed to: -Review their inventory and return the affected part(s) using the provided label to return the inventory in their possession. Any questions contact Dalton Daggs at Dalton.daggs@seaspine.com or Danielle Nelson-Archer at Danielle.nelson-archer@seaspine.com.
- Initiated
- 2024-10-29
- Root cause
- Nonconforming Material/Component
- Distribution
- U.S. Nationwide distribution in the states of CA, CO, FL, GA, PA, TX, and VA.