Open original FDA record ↗Recall record
- FDA identifier
- Z-0752-2016
- Company in FDA record
- Toshiba American Medical Systems Inc
- FDA classification date
- Feb 9, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
When a fontal DA (Digital Angiography) acquisition was done, scattered x-ray came into the dose meter on the lateral side. As a result, the dose meter sent a minus value to the software. Consequently, the software defined the data as an "abnormal value" and it stopped displaying dose data and the dose data was lost. The following message was displayed, "Dose meter abnormal, Dose info disabled".
Source facts
- Action
- Toshiba's planned action to bring into compliance: 1.Affected customers will be notified using the Customer Notification Letter you provided, 2. revised system software will be installed on the systems to prevent this occurrence from happening, 3. the corrections will be made free of charge, and 4. the corrections will be completed by April 4, 2016. For further questions, please call 1- 800 521-1968.
- Initiated
- 2015-08-14
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2016-05-18
- Distribution
- US Nationwide Distribution to OH and NY
- Recall number
- Z-0752-2016
- Recall status
- Terminated
- CDRH record ID
- 143016
- FDA classified
- 2016-02-09
- Recall event number
- 73078