Skip to content

Public FDA record

Z-0752-2017: Metzenbaum Scissors 7, Curved, Sterile The Metzenbaum Scissors 7, Curved Curved are single-use sterile s

Recall record Z-0752-2017 for Integra York PA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0752-2017
Company in FDA record
Integra York PA, Inc
FDA classification date
Dec 13, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Chevron seals of 6x10 Tyvek peel pouch used to package the sterile instruments may separate causing the packages to open and compromise sterility of the instrument.

Source facts

Action
The firm, Integra, mailed an "URGENT MEDICAL DEVICE RECALL NOTICE" dated October 26, 2016 to customers affected by the recall to notify them of the issue. The notice described the product, problem and actions to be taken. The customers were instructed to immediately review their inventory to determine if they had any of the recalled product on hand and to discontinue distributing the product; review your customer traceability records for shipment of the product and forward a copy of this Recall Notice to any of your customers that purchased the affected lots; complete and return the Acknowledgement Form that accompanied the notice (even if you do not have any product(s)) by email to: FCA3@integralife.com or fax to: 1-609-750-4220. Should you have any questions regarding these instructions, please contact Customer Service at 1-866-854-8300.
Initiated
2016-10-28
Root cause
Process control
Terminated
2017-04-17
Distribution
US Distribution in NC and TX.
Recall number
Z-0752-2017
Recall status
Terminated
CDRH record ID
151084
FDA classified
2016-12-13
Recall event number
75710

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗