During the subsequent pipetting (1st or 2nd measurement), the residual reagent can sporadically be transferred from the conus into the reaction cuvette and lead to inaccurately low results for the affected TDMs (Therapeutic Drug Monitoring).
Source facts
Action
On 11/20/2020, Roche issued Urgent Medical Device Correction notices to customers via phone and fax. Customers were advised to do the following:
A Roche Customer Support Center representative will contact the four affected customers by phone and a copy of UMDC TP-01155 will be faxed to the customer. Any customer questions or concerns will be addressed during the phone call.
Initiated
2020-11-23
Root cause
Device Design
Terminated
2023-07-06
Distribution
Domestic: IL, FL, GA, WV
Recall number
Z-0752-2021
Recall status
Terminated
CDRH record ID
184818
FDA record created
2020-12-29
Recall event number
86897