Skip to content

Public FDA record

Z-0752-2025: Brand Name: Daytona Small Stature Product Name: Ti Locking Cap 4.5mm, Small Stature, T25 Model/Catalog Nu

Recall record Z-0752-2025 for SEASPINE ORTHOPEDICS CORPORATION. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0752-2025
Company in FDA record
SEASPINE ORTHOPEDICS CORPORATION
FDA classification date
Dec 23, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to manufacturing defects observed on Locking Caps that affect the ability for components to mate.

Source facts

Action
Update 12/18/2024 - The firm sent an updated Customer Notification Letter on 12/12/2024 to customers that have not responded to the initial notification on 10/29/2024. On 10/29/2024, firm sent an "URGENT: MEDICAL DEVICE RECALL" Letter to customer informing them that the firm has identified a nonconformance with the threads of the locking cap components that may be incongruous with optimal manufacturing specifications. The nonconformance can result in a locking cap that does not mate appropriately with the screw tulip. Customers are instructed to: -Review their inventory and return the affected part(s) using the provided label to return the inventory in their possession. Any questions contact Dalton Daggs at Dalton.daggs@seaspine.com or Danielle Nelson-Archer at Danielle.nelson-archer@seaspine.com.
Initiated
2024-10-29
Root cause
Nonconforming Material/Component
Distribution
U.S. Nationwide distribution in the states of CA, CO, FL, GA, PA, TX, and VA.
Recall number
Z-0752-2025
Recall status
Open, Classified
CDRH record ID
211127
FDA classified
2024-12-23
Recall event number
95760

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗