Skip to content

Public FDA record

Z-0753-2016: GE Precision MPi is an all-digital multipurpose tilt-C x-ray system, intended for a multitude of diagnost

Recall record Z-0753-2016 for Regulatory Insight, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0753-2016
Company in FDA record
Regulatory Insight, Inc
FDA classification date
Feb 24, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It was discovered that the Remote Touch Panel (RTP) of the GE Precision MPi X-ray system may not always boot up as intended and needs to be updated to properly accomplish its intended purpose.

Source facts

Action
Nordisk Rontgen Teknik ( NRT ) Planned action to bring defect into compliance: 1.Nordisk Rontgen Teknik A/S (NRT) will contact customers and initiate a software and hardware update to correct the defect. 2.NRT engineers will install the software and perform testing to ensure the software and hardware update was effective. 3.The customer notification letter which includes a statement that NRT will without charge, remedy the defect or bring the product into compliance. For further questions please call (720) 962- 5412.
Initiated
2008-07-01
Root cause
Radiation Control for Health and Safety Act
Terminated
2017-07-11
Distribution
US Distribution to the states of :TN, CA, WA, OR, PA, WI, FL, MN, VT, NY, NV, MA, CO, IN, MI, MS and OH.
Recall number
Z-0753-2016
Recall status
Terminated
CDRH record ID
141343
FDA classified
2016-02-24
Recall event number
72515

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗