Z-0753-2020: Medicrea Pass LP, REF B02217545, Non Sterile, Material: Ti-6Al-4V ELI, VIS PEDICULAIRE POLYAXIALE 7.5MM x

Recall record Z-0753-2020 for Medicrea International. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0753-2020
Company in FDA record
Medicrea International
FDA record creation date
Dec 30, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It has been determined that some Pedicle system screw identified as 45mm long have an actual length of 50mm. Pedicle screws with a diameter of 7.5 mm or more are reserved for the sacrum area. As a result, a screw that is too long could come into contact with the nerves and cause postoperative pain.

Source facts

Action
On June 27, 2019, Medicrea begin emailing a "field safety correction action - Recall" notification to affected consignees. In addition to informing the consignee about the recall, the notification asked customers to do the following: 1. Immediately check your internal inventory and quarantine all subject devices pending return to MEDICREA 2. Circulate this Field Safety Notice internally to all interested/affected parties. 3. Inform MEDICREA if any of the subject devices have been distributed to other organisations. a) Please provide contact details so that MEDICREA can inform the recipients appropriately. b) If you are a Distributor, note that you are responsible for notifying your affected customers. 4.. Please inform MEDICREA of any adverse events concerning the use of the subject devices Please comply with any local laws or regulations concerning the notification of adverse events to your National Competent Authority. 5. Complete the attached customer response form. It may be that you no longer have any physical inventory on site. Completing this form will allow us to update our records and will also negate the need for us to send any further unnecessary communications on this matter. Therefore, please complete even if you no longer have any of the subject devices in your physical inventory. 6. We request that you respond to this notice within 07 calendar days from the date of receipt.
Initiated
2019-06-27
Root cause
Employee error
Terminated
2020-11-19
Distribution
US: NY, CO, OH, AZ, FL, TX, PA OUS: France, Belgium, South Africa, China, Spain, Portugal
Recall number
Z-0753-2020
Recall status
Terminated
CDRH record ID
177938
FDA record created
2019-12-30
Recall event number
84273

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗