Open original FDA record ↗Recall record
- FDA identifier
- Z-0755-05
- Company in FDA record
- Ethicon Endo-Surgery
- FDA classification date
- Apr 27, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Due to tolerance issues with the device, pins can work free from their respective holes. The result during use could be separation of the handles that could lead to compromised ratchet button functionality. Loss of ratchet button functionality may lead to the end effector being locked in the open or closed position.
Source facts
- Action
- The recalling firm sent letters to hospitals and distributors, dated 12/14/04.
- Initiated
- 2004-12-15
- Root cause
- Other
- Terminated
- 2012-05-30
- Distribution
- The devices were distributed to hospitals and distributors located throughout the United States and to foreign consignees located in Saudi Arabia, Austria, Egypt, United Kingdom, Germany, France, Lebanon, The Netherlands, Spain, Pakistan, South Africa, Italy, United Arab Emirate, Czech Republic, Sweden, Switzerland, Greece, India, Portugal, Slovenia, Hungary, Palestine, and Kuwait.
- Recall number
- Z-0755-05
- Recall status
- Terminated
- CDRH record ID
- 36863
- FDA classified
- 2005-04-27
- Recall event number
- 30727