Open original FDA record ↗Recall record
- FDA identifier
- Z-0757-2017
- Company in FDA record
- Pentax of America Inc
- FDA record creation date
- Dec 14, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The EB-1970UK Ultrasound Video Bronchoscope shipped between the clearance date of April 2014 and September 2015 contained the Operation and Reprocessing IFUs that were in use prior to the 2014 clearance, and therefore did not include the cleared Operation and Reprocessing IFUs.
Source facts
- Action
- Pentax sent a US Urgent Field Correction letter to all affected customers with a field correction/field safety response form via USPS on May 31, 2016. The letter identified the product, the problem, and the action to be taken by the customer. Revised instructions were as follows: The instructions have been revised in the following manner: Operation IFU o Integrated all models of each family into the IFU o Modified the indications for use language to be consistent with similar products o Enhanced the explanation of the operation of the device o Revised and added cautions and warnings Reprocessing IFU o Included more specificity regarding reprocessing activities to aid in the proper method for reprocessing the endoscopes, including: number of brush and rinse steps, quantity of fluid volumes, use of additional text, figures, cautions, and warnings Identification of Affected Endoscope Models Table 1 provides an overview of the Affected Endoscope models, and the part number and revision of the historical and revised IFUs. The accompanying USB flash drive contains all the revised IFUs identified below. Please distribute only those IFUs which support the products owned by your facility. Table 1 Affected Models Historical Version of IFUs Revised IFUs EB-1970UK Ultrasound Video Bronchoscope Z870 Owners IFU S016_R02 Operator IFU S021_R04 Reprocessing IFU No.388_R01 Addendum No.392_R00 Addendum EB-70K series Video Bronchoscopes Z705 Operation IFU S015_R04 Operator IFU S020_R04 Reprocessing IFU No.388_R01 Addendum No.392_R00 Addendum Z706 Reprocessing IFU EB-75K series Video Bronchoscopes S003 Owners IFU Customers were instructed to ensure that user departments discard the prior versions of the Instructions for Use and ensure that all users read and carefully follow the revised Operation and Reprocessing IFUs to ensure the proper use and cleaning of the affected PENTAX endoscopes. Customers were advised to contact their PEN
- Initiated
- 2016-05-31
- Root cause
- Process control
- Terminated
- 2019-08-19
- Distribution
- Nationwide Distribution
- Recall number
- Z-0757-2017
- Recall status
- Terminated
- CDRH record ID
- 149791
- FDA record created
- 2016-12-14
- Recall event number
- 75249