Z-0758-04: Baxter Meridian Hemodialysis Instrument, product codes 5M5576 and 5M5576R; Baxter Healthcare Corporation,

Recall record Z-0758-04 for Baxter Healthcare Renal Div. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0758-04
Company in FDA record
Baxter Healthcare Renal Div
FDA classification date
Jul 20, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Microbubbles of air in the extracorporeal blood circuit during hemodialysis therapy on the Meridian.

Source facts

Action
Urgent Device Correction letters dated 3/11/04 were sent to all Meridian users, informing them of adverse events associated with air in the extracorporeal blood circuit during hemodialysis therapy on the Meridian. The letter outlined the clinical practices that must be followed for all hemodialysis therapies to reduce the potential for air to be introduced into the extracorporeal blood circuit during hemodialysis. Any questions or related events to report were directed to Baxter Instrument Services at 1-800-553-6898, select prompt 3, option 2 and choice 2. Baxter developed a permanent fix -- an enhanced air detector with improved sensitivity, making it capable of detecting air bubbles of a smaller size. Letters were sent to all Meridian customers on 7/11/05, informing them of the availability of the new detector and that a Baxter Field Service Engineer would contact them to schedule the upgrade.
Initiated
2004-03-10
Root cause
Other
Terminated
2006-10-26
Distribution
Nationwide, Mexico, China and Korea.
Recall number
Z-0758-04
Recall status
Terminated
CDRH record ID
31970
FDA classified
2004-07-20
Recall event number
28475

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗