Open original FDA record ↗Recall record
- FDA identifier
- Z-0758-2021
- Company in FDA record
- Boston Scientific Corporation
- FDA record creation date
- Dec 30, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Trace amounts of metallic debris were found on the stent related to the manufacturing process. The debris was confirmed to be particles of nickel.
Source facts
- Action
- On November 16, 2020, the firm distributed Urgent Medical Device Removal letters to affected customers. Customers were informed of the product issue and asked to take the following actions: - Verify whether you have any product from the affected lots/batches within your inventory - Immediately cease use or distribution of any remaining unused product from the affected lots/batches - Segregate affected material and return it to BSC in accordance with removal instructions included in the letter. - Distributors should forward the notification to their customers Your local Sales Representative can answer any questions that you may have regarding this notification.
- Initiated
- 2020-11-16
- Root cause
- Process design
- Distribution
- Distribution US nationwide and Japan.
- Recall number
- Z-0758-2021
- Recall status
- Open, Classified
- CDRH record ID
- 184742
- FDA record created
- 2020-12-30
- Recall event number
- 86880