Skip to content

Public FDA record

Z-0759-06: The unimplanted HiRes90K Implantable Cochlear Stimulators (ICS) manufactured with feedthrus from a speci

Recall record Z-0759-06 for Advanced Bionics Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0759-06
Company in FDA record
Advanced Bionics Corporation
FDA classification date
Apr 11, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Certain HiRes 90K cochlear implants could fail as a result of elevated moisture levels. The unimplanted HiRes 90K implants listed are being removed from the market. Patients are to be followed for signs of device failure: intermittent function, complete loss of sound, sudden discomfort, pain, noise, or popping.

Source facts

Action
The recall was initiated on 03/08/06. In North America, the clinician notification letters were distributed by fax. For other regions, an Advanced Bionics representative and/or distributor are providing notification to the affected clinicians. As part of the clinician notification process, each center was provided with a list of their affected devices. Customers were given a list of HiRes 90K implants with feedthrus from Supplier B that were shipped to their clinic and are not yet registered with Advanced Bionics. Customers were asked to return the implants listed in the attachment to Advanced Bionics according to the instructions. Customers were told not to implant any of the devices on the list. Customers were told that implants manufactured wth feedthrus from Supplier A will be delivered quickly to minimize disruptions to their schedule. Customers will be contacted through Advanced Bionics Customer Service within 48 hours to arrange retrieval of implants manufactured with feedthrus from Supplier B and replacement with implants manufactured with feedthrus from Supplier A.
Initiated
2006-03-08
Root cause
Other
Terminated
2012-03-23
Distribution
Nationwide to: AR, AL, FL, TX, NY, NE, NC, CO, GA, OH, IL, PA, IN, MN, NC, CA, HI, MA, AZ, VA, OR, SD, MI, MO, IA, UT, TN. Worldwide to: Argentina, Austria, Belgium, Brazil, Canada, Chile, China, Columbia, Costa Rica, Cuba, Denmark, Ecuador, Egypt, France, Germany, Hong Kong, India, Israel, Italy, Japan, Lebanon, Mexico, Morocco, Netherlands, Norway, Poland, Portugal, Puerto Rico, Serbia, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, Venezuela.
Recall number
Z-0759-06
Recall status
Terminated
CDRH record ID
44983
FDA classified
2006-04-11
Recall event number
34873

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗