Open original FDA record ↗Recall record
- FDA identifier
- Z-0759-2019
- Company in FDA record
- Stihler Electronic Gmbh
- FDA record creation date
- Jan 23, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
PRISMAFLO II and PRISMAFLO II S Heating Profile only and not the Control Unit, damage may result in a hotspot on the Heating Profile, may cause low level skin burn
Source facts
- Action
- STIHLER ELECTRONIC GMBH notified distributors by Registered mail/email and to Head of Biomedical on 09/06/18 identifying the problem, health risk and action to be taken: identify the Heating Profiles according to the serial numbers specified. These profiles must be replaced based on your accurate reply. Complete Inventory must be accounted for. For your reply use attached "Return Form" to Stihler Electronic GmbH. Send the filled return form via e-mail to nicola. vosseler@sti hlerelectronic. de. 3. You will receive the replacement Heating Profiles from our distributor Futuremed America Inc., 15700 Devonshire Street, Granada Hills, 91344 California, US.
- Initiated
- 2018-09-06
- Root cause
- Device Design
- Terminated
- 2020-05-13
- Distribution
- Worldwide Distribution - US Nationwide in the states of CA and Foreign of: Switzerland
- Recall number
- Z-0759-2019
- Recall status
- Terminated
- CDRH record ID
- 169729
- FDA record created
- 2019-01-23
- Recall event number
- 81755