Open original FDA record ↗Recall record
- FDA identifier
- Z-0760-2007
- Company in FDA record
- Nellcor Puritan Bennett
- FDA classification date
- Apr 24, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
In rare instances, the safety valve may remain open after an occlusion is resolved. If the safety valve does not reset, the ventilator will detect a circuit disconnect condition. When this happens, the ventilator will annunciate a high-urgency audible alarm and the ventilator will not resume normal operation.
Source facts
- Action
- The correction will extend to the end user. All users will be contacted by mail or through personal contact. In the U.S., all customers were mailed the 'Urgent Device Correction' letter along with a serial number listing of those units which records show were consigned to them beginning on March 12, 2007. The customer communication informs customers of the issue and provides instruction for continued safe use of the device pending the correction.
- Initiated
- 2007-03-12
- Root cause
- Other
- Terminated
- 2012-12-13
- Distribution
- Worldwide, including USA, ARGENTINA, AUSTRALIA, BELGIUM, BRAZIL, CANADA, CHILE, COLOMBIA, EGYPT, FRANCE, GERMANY, GREECE, GUATEMALA, IRAN, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, KOREA, LEBANON, MEXICO, MOROCCO, NICARAGUA, PANAMA, POLAND, PUERTO RICO, ROMANIA, RUSSIA, SINGAPORE, SOUTH AFRICA, SPAIN, TURKEY, UK, & YUGOSLAVIA
- Recall number
- Z-0760-2007
- Recall status
- Terminated
- CDRH record ID
- 51015
- FDA classified
- 2007-04-24
- Recall event number
- 37607