A medical device used in brachytherapy may become contaminated and can cause infections, toxicity or allergic reactions.
Source facts
Action
Varian Medical Systems an an Urgent Medical Device Correction letters dated April 11, 2012 and May 8, 2012 to all affected customers. The April notification advised of an anomaly with titanium rectal retractors, and the correction was expanded in May to include stainless steel retractors. In both notifications, users were instructed to discontinue use and quarantine their cleaned and sterilized devices for collection by Varian field service engineers. Users were requested to return an enclosed reply form to respond with number of units in inventory.
Initiated
2012-04-12
Root cause
Component design/selection
Terminated
2014-04-10
Distribution
US Nationwide Distribution
Recall number
Z-0760-2013
Recall status
Terminated
CDRH record ID
114746
FDA classified
2013-01-31
Recall event number
61879