Open original FDA record ↗Recall record
- FDA identifier
- Z-0760-2016
- Company in FDA record
- C.R. Bard, Inc.
- FDA classification date
- Feb 5, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Improper targeted temperature therapy was delivered to patients because the Temperature Simulator Key was connected to the patient temperature cable instead of the patient temperature probe.
Source facts
- Action
- Bard Medical Division sent a Safety Alert to each of the customers via FedEx with proof of delivery notice. All consignees have been instructed to carry the notification out to the user-level.
- Initiated
- 2015-11-10
- Root cause
- Use error
- Terminated
- 2017-04-21
- Distribution
- Nationwide Distribution-DC and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI. IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV.
- Recall number
- Z-0760-2016
- Recall status
- Terminated
- CDRH record ID
- 142163
- FDA classified
- 2016-02-05
- Recall event number
- 72756