Open original FDA record ↗Recall record
- FDA identifier
- Z-0760-2023
- Company in FDA record
- Baxter Healthcare Corporation
- FDA classification date
- Dec 22, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for Internal Blood Leaks due to a twisted gasket in the dialyzer
Source facts
- Action
- Baxter Healthcare sent a notification via USPS first class mail on about 11/23/2022 and to home patients on about 11/29/2022. Consignees were instructed to locate and remove all affected product from your inventory, contact Baxter Healthcare Center for Service to arrange for return and credit, and acknowledge the receipt of the notification by completing a reply form on the customer portal. Note: this letter describes two issues. This recall addresses Internal Blood Leaks being caused by a twisted gasket in the dialyzer. The second issue is linked to reports of Polyurethane Displacement in the dialyzer, which is addressed in a separate recall.
- Initiated
- 2022-11-23
- Root cause
- Process change control
- Distribution
- Worldwide distribution. US Nationwide including Guam and St. Thomas USVI; Canada, China, Australia, New Zealand, Austria, Egypt, France, Germany, Ireland, Italy, Morocco, Qatar, Saudi Arabia, South Africa, Spain, United Kingdom, Bahamas, Barbados, Belize, Bermuda, Colombia, Dominican Republic, Ecuador, Guatemala, Honduras, Mexico, Panama, and Trinidad and Tobago.
- Recall number
- Z-0760-2023
- Recall status
- Open, Classified
- CDRH record ID
- 197272
- FDA classified
- 2022-12-22
- Recall event number
- 91222