Open original FDA record ↗Recall record
- FDA identifier
- Z-0761-2007
- Company in FDA record
- Bio-Logic Systems Corp
- FDA classification date
- Apr 20, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Battery and Battery Compartment; incorrect placement in the Netlink Traveler unit which caused a short in the wiring of the battery pack and overheating.
Source facts
- Action
- On 2/26/07, Bio-logic Systems shipped replacement batteries, marked with part number 520-LIBAT4-3 and a label reading 'THIS SIDE OUT', and a new battery compartment cover with a label inside the cover to clarify the correct installation of the battery to each of the affected customers. The accounts were requested to remove and replace the old batteries, part 520-LIBTA4-2, and compartment covers, and return the old batteries and covers to Bio-logic Systems in the enclosed pre-paid shipping container, along with the completed accounting form.
- Initiated
- 2007-02-26
- Root cause
- Other
- Terminated
- 2008-08-26
- Distribution
- Worldwide, USA , Canada, China, Hong Kong, India and Israel.
- Recall number
- Z-0761-2007
- Recall status
- Terminated
- CDRH record ID
- 50670
- FDA classified
- 2007-04-20
- Recall event number
- 37518